Owner manual

epBlue - Software error messages
epBlue™ with MultiCon
English (EN)
170
0x2122 Parameter conflict:
mixing volume in
source too small for
this tool
The Volume and Pipet. Tool
parameters of a mix procedure
for source vessels do not
match (Volume is less than the
lower limit of the tool's
volume range) (mix procedure
as defined in a command Mix
or as part of a liquid transfer
command via parameter
Options).
Edit Volume and Pipet. Tool
so that Volume is within
the pipette tool's volume
range.
0x2123 Parameter conflict:
mixing volume in
source greater than
filling volume of
source tube or well
The Volume and Source
parameters of a mix procedure
in the application do not
match (Volume is higher than
the maximum filling volume of
the source vessel) (mix
procedure as defined in a
command Mix or as part of a
liquid transfer command via
parameter Options).
Edit Volume and Source so
that Volume is within the
allowable filling volume of
the source vessel.
0x2124 Mixing cycles in
destination not
defined
Parameter No. of Cycles in a
mix procedure for destination
vessels was not edited in the
application (mix procedure as
part of a liquid transfer
command via parameter
Options)
See explanation in
"Cause".
0x2125 Invalid entry for
mixing cycles in
destination
(1 ... 99)
Entry for the parameter No. of
Cycles in a mix procedure for
destination vessels was higher
than the max. limit (1 up to
99 cycles) (mix procedure as
part of a liquid transfer
command via parameter
Options).
Enter a number between
1 and 99 for the No. of
Cycles parameter.
0x2126 Invalid entry for
mixing speed in
destination
(1 ... 10)
Entry for the parameter Speed
in a mix procedure for
destination vessels was higher
than the max. limit (1 up
to 10) (mix procedure as part
of a liquid transfer command
via parameter Options).
Enter a number between
1 and 10 for the Speed
parameter.
Code Symptom/message Cause Remedy