Owner manual
epBlue - Software error messages
epBlue™ with MultiCon
English (EN)
167
0x2026 Bottom tolerance
too small
Bottom tolerance too small. Use a bigger value.
0x2101 Tool not defined Parameter Pipet. Tool was not
edited in the application.
See explanation in
"Cause".
0x2104 Tips not edited in
worktable/
procedure
Tips were edited in the
procedure of the application,
but they were not edited in the
worktable (e.g., filter tips <->
tips without filter).
Edit the tips that you
programmed in the
procedure in the
worktable.
0x2105 Parameter conflict:
Start volume
greater than filling
volume of source
tube or well
The Volume and Source
parameters of the source
vessel do not match (Volume is
higher than the maximum
filling volume of the source
vessel).
Edit Volume and Source in
the worktable so that
Volume is covered by the
maximum filling volume of
the source vessel.
0x2106 Parameter conflict:
Start volume
greater than filling
volume of
destination tube or
well
The Volume and Destination
parameters of the destination
vessel do not match (Volume is
higher than the maximum
filling volume of the
destination vessel).
Edit Volume and
Destination in the
worktable so that Volume is
covered by the maximum
filling volume of the
destination vessel.
0x2107 Volume not defined Parameter Volume was not
edited in the application.
See explanation in
"Cause".
0x2108 Parameter conflict:
Volume too small
for this tool
The Volume and Pipet. Tool
parameters of the application
do not match (Volume is
smaller than the lower limit of
the tool volume range).
Edit Volume and Pipet. Tool
so that Volume is covered
by the volume range of the
pipette tool.
0x2109 Parameter conflict:
Volume too large
for this tool
The Volume and Pipet. Tool
parameters of the application
do not match (Volume is
higher than the upper limit of
the tool volume range).
Edit Volume and Pipet. Tool
so that Volume is covered
by the volume range of the
pipette tool.
0x210A Parameter conflict:
volume greater
than filling volume
of source tube or
well
The Volume and Source
parameters of the application
do not match (Volume is
higher than the maximum
filling volume of the source
vessel).
Edit Volume and Source so
that Volume is covered by
the maximum filling
volume of the source
vessel.
0x210B Parameter conflict:
volume greater
than filling volume
of destination tube
or well
The Volume and Destination
parameters of the application
do not match (Volume is
higher than the maximum
filling volume of the
destination vessel).
Edit Volume and
Destination so that Volume
is covered by the
maximum filling volume of
the destination vessel.
Code Symptom/message Cause Remedy