Owner manual
epBlue - Software error messages
epBlue™ with MultiCon
English (EN)
165
0x1C15 Pattern too
complex
Edit mode/pattern:
The pattern algorithm cannot
handle this pattern.
Note: See note in error
0x1C0F.
Enter a simpler pattern if
possible.
In case this is not possible:
Call local Eppendorf
Service.
0x1C16 This position is
already occupied
Edit mode/pattern:
When editing the pattern you
have tried to select a certain
rack position that is already
occupied.
Following the edited
pattern move to a different
rack position.
0x1C17 You must start with
the source
Edit mode/pattern:
When editing a pattern you
must start with the source.
See explanation in
"Cause".
0x1C18 Please enter a
source now
Edit mode/pattern:
In the destination rack, you
tried to enter more replicates
than you had sources.
Enter the same number of
replicates for all sources
you edit when
programming a pattern.
0x1C19 Please enter a
destination now
Edit mode/pattern:
When having selected a
source in the Sample Transfer
command you first have to
enter a destination for this
source before moving to the
next source position.
Press the Destination
softkey and edit the
destination position(s) for
the selected source
position.
0x1C1A No more positions
available (limited
by Number of
Samples
command)
Edit mode/pattern:
Editing further positions is not
possible because the limit set
in the Number of Samples
command would be exceeded.
Select a pattern that fits the
programmed Number of
Samples command.
0x1C1E Pattern for reagent
transfer:
Source can be
chosen only once
Edit mode/pattern for
command Reagent Transfer:
After having entered the
source and the destinations
for the reagent transfer you
cannot select an additional
source.
Enter the source only once.
In case this does not meet
your requirements for this
application consider
selecting command Sample
Transfer instead of Reagent
Transfer, or:
Call local Eppendorf
Application Support.
0x1C1F Pattern with
8-channel tool:
Please edit upper
position of this tool
Edit mode/pattern with
8-channel tool:
Only the upper positions of the
8-channel tools can be
selected.
See explanation in
"Cause".
Code Symptom/message Cause Remedy