Owner manual
epBlue - Software error messages
epBlue™ with MultiCon
English (EN)
161
0x1A22, 0x1A32,
0x1A42, 0x1A52
No source or
destination selected
Edit mode/parameter in the
affected command:
Source or destination rack has
not been selected.
Select source or
destination, respectively.
0x1A40, 0x1A41,
0x1A50
8-channel tool
cannot be used for
this source rack
Edit mode/parameter in the
affected command:
Source rack does not fit the
8-channel tool (e.g., 24-well
plate or tube rack with
24 positions).
Choose another rack or
another tool.
0x1A61 8-channel tool
cannot be used for
this rack
Edit mode/parameter in
command Mix:
Rack does not fit the
8-channel tool (e.g., 24-well
plate or tube rack with
24 positions).
Choose another rack or
another tool.
0x1A62 No rack selected Edit mode/parameter in
command Mix:
Rack has not been selected.
Select rack, respectively.
0x1A65 Invalid entry for
speed (0.2 ... 110)
Edit mode/parameter in
command Mix:
A value beyond the allowed
range has been entered for the
parameter Speed.
Enter a value between
0.2 and 110 mm/sec.
0x1A70 This position is
already occupied
Edit mode/pattern:
When editing the pattern you
have tried to select a certain
position that is already
occupied.
Follow the direction of the
edited pattern and move to
a different position.
0x1A76 8-channel tool
cannot be used for
this module rack
Edit mode/pattern:
Rack does not fit the
8-channel tool (e.g., Tubs +
Modules (equip) + Holders
combination with positions all
less than 8 in modules).
Choose another rack or
another tool.
0x1A77 No module rack or
tubes found
Edit mode/pattern:
Rack does not have any
positions (e.g., Tubs + Modules
(equip) + Holders-combination
with positions all less than 1 in
Modules).
Choose another rack.
Code Symptom/message Cause Remedy