Owner manual

epBlue - Software error messages
epBlue™ with MultiCon
English (EN)
159
0x1A10 8-channel tool
cannot be used for
this source rack
Edit mode/parameter in
command Sample Transfer:
Source rack does not fit to
8-channel-tool (e.g.:
24-well-plate or tube rack with
24 positions).
Choose another rack or
another tool.
0x1A11 8-channel tool
cannot be used for
this destination
rack.
Edit mode/parameter in
command Sample Transfer:
Destination rack does not fit
the 8-channel tool (e.g.,
24-well plate or tube rack with
24 positions).
Choose another rack or
another tool.
0x1A12 No source or
destination selected
Edit mode/parameter in
command Sample Transfer:
Source or destination rack has
not been selected.
Select source or
destination, respectively.
0x1A15 Invalid entry for
movement blow
(0 ... 100)
Edit mode/parameter in
transfer command:
A value beyond the allowed
range has been entered for the
parameter Movement Blow.
Enter a value between
0 and 100%.
0x1A15 Invalid entry for
movement blow
(0 ... 100)
Edit mode/parameter in
transfer command:
A value beyond the allowed
range has been entered for the
parameter Movement Blow.
Enter a value between
0 and 100%.
0x1A16 Invalid entry for
delay blow
(0 ... 9999)
Edit mode/parameter in
transfer command:
A value beyond the allowed
range has been entered for the
parameter Delay Blow.
Enter a value between
0 and 9999 msec.
0x1A16 Invalid entry for
delay blow
(0 ... 9999)
Edit mode/parameter in
transfer command:
A value beyond the allowed
range has been entered for the
Delay Blow parameter.
Enter a value between
0 and 9999 msec.
0x1A17 Invalid entry for
speed aspiration
(0.2 ... 110)
Edit mode/parameter in
transfer command:
A value beyond the allowed
range has been entered for the
parameter Speed Aspiration.
Enter a value between
0.2 and 110 mm/sec.
Code Symptom/message Cause Remedy