0x12D1 Parameter conflict:
Elution volume too
large for
destination tube or
well
Option elution from filter:
volume is too large for the
vessel used.
Select a tool large enough
when using this option.
0x12E1 Parameter conflict:
Pre-wetting not
possible when
aspirate from
bottom is selected
See error message.
Change application.
0x12E2 Parameter conflict:
Pre-wetting not
possible when
dispense from top
is selected
See error message.
Change application.
0x12E3 Parameter conflict:
Pre-wetting not
possible when
elution from filter is
selected
A liquid type using a
pre-wetting step (e.g., ethanol
98%) cannot be used in
combination with the elution
from filter parameter in a
sample transfer command.
Change application.
0x12E4 Transport of
labware: danger of
collision
Danger of collision when
transporting a stack of several
racks. The upper rack would
collide with the carrier.
Do not stack more than 5
racks in a location.
0x12E6 Level too high The liquid level would be
higher than the vessel after
dispensing.
Adjust the liquid to be
dispensed to the vessel.
0x12E7 Opening the hood
is not allowed when
putting down tool.
Switch off power,
then switch on
again to restart
method.
See error message.
See error message.
0x12F3 Thermomixer is too
hot for labware
(temperature
command in this
method)
The temperature unit of the
thermomixer is to hot for the
selected labware.
Choose a lower temperature.
0x12F4 The thermomixer is
too hot for labware
(temperature
command in
previous method)
The temperature unit of the
thermomixer is to hot for the
selected labware.
Choose a lower temperature.
Code Symptom/message Cause Remedy