0x128E Carrier: steps lost
in y
Carrier collision.
Carrier mechanics out of
order.
Ensure that the worktable
equipment in epBlue is
identical to the equipment
in epMotion, then restart
the application.
If the error persists, call
Eppendorf Service.
0x1294 Carrier: steps lost
in z
Carrier collision.
Carrier mechanics out of
order.
Ensure that the worktable
equipment in epBlue is
identical to the equipment
in epMotion, then restart
the application.
If the error persists, call
Eppendorf Service.
0x1295 Carrier:
steps lost in z
before picking up
tip
Tip was still on pipette tool
when tool started to pick
up a new tip.
Tip rack not placed
correctly on the worktable.
Mechanical problems of
carrier.
Remove tips from tools.
Place tip rack correctly,
and plane, on the
worktable.
In other cases:
Call local Eppendorf
Service.
0x1296 Maximum number
of tool cycles
exceeded
See error message.
Have the tool serviced.
0x1297 Danger of collision When running the
programmed application, the
tool carrier system will touch
racks or other labware on the
worktable; e.g., during
pipetting, the optical sensor
may touch a long tube on the
adjacent position; possible
reasons:
A low plate (microplate) is
located next to a high tube
rack.
The 50 μL or 300 μL tip is
programmed to move
almost to the bottom of a
very long tube with
another long tube in the
adjacent position.
Program the labware on
the worktable so that high
and low labware are not
adjacent.
Program the labware so
that the 30 µL or 300 µL tip
does not have to move
deeply into a long vessel.
If possible: use higher
volumes in the long
vessels.
If possible: use longer tips
for the long vessels.
0x1298 Tool not calibrated Tool not calibrated. Have the tool serviced.
0x1299 Invalid number of
samples
Value for number of samples
not permissible.
Insert an admissible value
for number of samples.
Code Symptom/message Cause Remedy