Owner manual
epBlue settings - Checking components and making the basic settings
epBlue™ with MultiCon
English (EN)
125
1. Define the speed and temperature in the Parameter area.
2. To check the mixing function, press the Start button.
The thermomixer mixes until the time is elapsed or until the test has been exited using the Stop button.
11.7.2 Checking the temperature control of the thermomixer
Abb. 11-7: Thermomixer temperature Control function
Fig. 11-7: Thermomixer temperature Control function
RPM field
Speed of the thermomixer in rpm.
Time field
Duration of the mixing process.
Remaining time field
Remaining time in the mixing process.
Start button
Stop button
Desired Temperature field
Set temperature of the thermomixer.
Actual Temperature field
Actual temperature of the thermomixer.
Actual current field
Current strength via the thermal module.
Time field
Mixing time in min.
Remain time field
Time until the end of the mixing process
(in min).
Temp reached after field
Time until the set temperature is reached.
New Temperature field
Enter a new set temperature. 4 °C– 95 °C range.
New time field
Enter a new mixing time. 0 s - 99 min range.
Start button
Start the test.
Fan on button
Switch on the fan.