Owner manual

epBlue settings - Checking components and making the basic settings
epBlue™ with MultiCon
English (EN)
124
11.7 Thermomixer
You can use this function to check the thermomixer.
Tab. 11-1: Maximum speed for labware on the thermomixer
11.7.1 Checking the mixing function of the thermomixer
Abb. 11-6: Mixer Control Thermomixer function
Fig. 11-6: Mixer Control Thermomixer function
NOTICE! Contamination of samples and the device due to high mixing frequency.
If the speed is set too high, liquid will spray out of the vessels. Labware may become loose
from the holder and fly about.
Observe the maximum speeds of the labware in the operating instructions.
Labware Maximum speed in rpm
Tubes 1000
DWP 96 1200
DWP 384 1200
MTP 6 1000
MTP 96 2000
MTP 384 2000
PCR 96 2000
PCR 384 2000
Reservoir Rack Cannot be mixed
Thermorack/Thermoblock 1000
Thermoadapter 96 1000
Thermoadapter 384 1000
Thermorack TMX 1300
PrepRack 1300