Owner manual
epBlue settings - Checking components and making the basic settings
epBlue™ with MultiCon
English (EN)
123
11.6 Temperature control unit X
The Temperature control unit X function allows you to check the epMotion thermal modules.
Abb. 11-5: Temperature control unit function
Fig. 11-5: Temperature control unit function
The current status will be displayed in the State area.
You can make settings in the Set Temperature area.
1. Enter the set temperature in the New Temperature field.
2. Enter the test duration in the New time field.
3. Click on the Start button. The thermal module heats or cools until the set temperature has been reached
or the set time has elapsed. If the set temperature is reached before the set time has elapsed, the
thermal module maintains the set temperature until the end of the test.
4. To update the display during the test, press the Refresh button.
5. To activate the fan, press the Fan On button.
Desired Temperature
Shows the set temperature of the thermal
module.
Actual Temperature
Shows the set temperature of the thermal
module.
Actual current
Shows the current strength in the thermal
module.
Time
Shows the temperature control duration in min.
Remain time
Shows the time remaining until the end of
temperature control (in min).
Temp reached after
Shows the time required to reach the set
temperature.
New Temperature
Set temperature (–5°C to 110°C)
New time
Temperature control duration (0 to 99 min)