6.4 Cleaning the pipette clamp
The parts of the pipette clamp can be replaced or cleaned as follows.
Requirement
Easypet is disassembled (see p. 15).
1. Replace the part if necessary.
2. Clean the part if necessary. Autoclave the aspirating cone and the pipette adapter at 121°C,
1 bar overpressure for 20 min if necessary. Then allow to dry at room temperature.
Aspirating cone
Can be wiped with a moist cloth
Can be disinfected with alcohol (ethanol,
propanol) or with a disinfectant containing
alcohol
Can be steam autoclaved multiple times
Pipette adapter
Can be rinsed with demineralized water
Can be steam autoclaved multiple times
Can be replaced
Membrane filter
Can be steam autoclaved once
Dispose of if contaminated
Cannot be cleaned
Can be replaced
Sealing
Can be replaced
Hint!
Alternatively, you can also use a membrane filter with a pore size of 0.2 μm, for
example, type Minisart SRP from Sartorius stedim.