7.1 General errors
Symptom/message Cause Remedy
Liquid drips out of the
pipette.
Pipette adapter and/or
membrane filter inserted
incorrectly.
Remove the pipette adapter
and membrane filter and
reinsert them.
Pipette not inserted far
enough.
Continue to carefully insert
the pipette.
Pipette adapter damaged.
Replace pipette adapter.
Pipette damaged.
Replace pipette.
•Seal damaged.
Replace seal.
Liquid column too high.
Speed too low.
Slowly press the aspirating
button down further.
Aspiration capacity
reduced.
Membrane filter wetted.
Replace membrane filter.
No functional period
despite fully charged
battery.
Battery is too old.
Replace battery.
Pipette loose. Pipette adapter damaged.
Replace pipette adapter.
Bubble formation in
the pipette during
liquid aspiration.
Too high speed.
The aspirating button may
only be pressed lightly.
Too high speed.
The aspirating button may
only be pressed lightly.
Rechargeable battery
will not charge.
Rechargeable battery is
charged.
Disconnect the mains/power
supply device.
Only charge the
rechargeable battery if the
status display is blinking.