11 Technical data
11.1 Power supply
11.2 Ambient conditions
Do not expose to direct sunlight.
11.3 Weight/dimensions
Power supply 100 V to 240 V ±10 %, 50 Hz to 60 Hz
Overvoltage category II
Degree of pollution 2
Power consumption Maximum power consumption according to name plate: 25 W
Approx. 15 W during operation
Approx. 5 W with the display dimmed
Permitted mains interruption Approx. 10 ms at 90 V
Approx. 20 ms at 230 V
Protection class I
Fuses T 2.5 A/250 V, 5 mm × 20 mm (2 pcs.)
Operation Ambient temperature: 15°C to 35°C
Rel. humidity: 25% to 70%
Air pressure: 86 kPa to 106 kPa
Air pressure Use up to an altitude of 2000 m above MSL
Weight 5.4 kg
Dimensions Width: 295 mm
Depth: 400 mm
Height:150 mm
Space required Width: 500 mm (with thermal printer: 750 mm)
Depth: 500 mm