Instruction Manual
17
Installation
Eppendorf BioPhotometer
®
D30
English (EN)
4.5 Connecting PC or USB stick for data export
You can connect a FAT 32-formatted USB stick to the USB port 4 (see Main illustration on p. 9).
Alternatively, you can connect the device for the data export directly to a PC by using a USB cable:
Prerequisites
• PC with Windows, version XP, SP2 or higher version.
• USB cable with a type A and type B plug each.
Connect the device to the PC by using the USB cable on the USB port 8 (see Main illustration on p. 9).
4 (ON) Zero = Normal
5 (ON) International
6 (ON) Character
7 (ON) Set
8 (OFF) = U.S.
DIP SW-3 Meaning
1 (ON) Data Length = 8 bits
2 (ON) Parity Setting = NO
3 (ON) Parity Condition = Odd
4 (OFF) Busy Control = XON/XOFF
5 (OFF) Baud
6 (ON) Rate
7 (ON) Select
8 (ON) = 9600 bps
• You do not need any special PC software for the data transmission: the transferred data
packets are recognized by the PC like a USB stick as a removable medium. For viewing the
data, you only need to open the registered data packet.
• The transmission of data to the USB stick or to the PC is started after completing the series
of measurement in the print & export (see Print & export on p. 43) method step.
DIP SW-2 Meaning