Design) User guide

Concentrator plus/Vacufuge® plus — Operating manual
40
9 Technical data
9 Technical data
9.1 Power supply
9.2 Ambient conditions
9.3 Weight / dimensions
5305 Basic device 5305 Complete system
Mains connection: 230 V, 50 to 60 Hz
120 V, 50 to 60 Hz
100 V, 50 to 60 Hz
230 V, 50 to 60 Hz
120 V, 50 to 60 Hz
100 V, 50 to 60 Hz
Current consumption: 2.1 A (230 V)
6.0 A (120 V)
5.2 A (100 V)
1.7 A (230 V)
5.6 A (120 V)
3.8 A (100 V)
Power consumption: max. 500 W (230 V)
max. 550 W (120 V)
max. 520 W (100 V)
max. 350 W (230 V)
max. 400 W (120 V)
max. 380 W (100 V)
Overvoltage category: II
Fuses: 4.0 AT (230 V)
6.3 AT (120 V/100 V)
Environment: For indoor use only.
Ambient temperature: 15 to 35 °C
Max. relative humidity: 75 %, non-condensing humidity
Atmospheric pressure: Use at altitudes up to 2000 m above MSL.
Degree of pollution: 2
5305 Basic device 5305 Complete system incl. gel
dryer
Dimensions: Width: 330 mm (12.6 in.)
Depth: 372 mm (14.5 in.)
Height: 231 mm (9.1 in.)
Width: 330 mm (12.6 in.)
Depth: 579 mm (20.9 in.)
Height: 290 mm (11.7 in.)
Weight without rotor: 16.5 kg (37.5 lbs.) 31.5 kg (68.3 lbs.)
Noise level: 50 dB(A) 50 dB(A)
EN
Operating manual