9.5 Acceleration times and deceleration times for the Centrifuge 5427 R
(according to DIN 58 970)
Rotor Acceleration time/Deceleration time Mains voltage
230 V 120 V 100 V
FA-45-12-17 Without soft ramp Acceleration time 27 s29 s29 s
Deceleration time 24 s25 s25 s
With soft ramp Acceleration time 39 s39 s39 s
Deceleration time 39 s39 s39 s
FA-45-24-11 Without soft ramp Acceleration time 18 s19 s19 s
Deceleration time 18 s19 s19 s
With soft ramp Acceleration time 29 s29 s29 s
Deceleration time 31 s31 s31 s
FA-45-24-11-Kit Without soft ramp Acceleration time 21 s22 s22 s
Deceleration time 21 s21 s21 s
With soft ramp Acceleration time 32 s32 s32 s
Deceleration time 31 s31 s31 s
FA-45-30-11 Without soft ramp Acceleration time 21 s22 s22 s
Deceleration time 18 s19 s19 s
With soft ramp Acceleration time 32 s32 s32 s
Deceleration time 33 s33 s33 s
F-45-30-11 Without soft ramp Acceleration time 21 s22 s22 s
Deceleration time 18 s19 s19 s
With soft ramp Acceleration time 29 s29 s31 s
Deceleration time 32 s32 s32 s
FA-45-48-11 Without soft ramp Acceleration time 28 s29 s29 s
Deceleration time 22 s23 s23 s
With soft ramp Acceleration time 36 s36 s39 s
Deceleration time 35 s35 s35 s
F-45-48-11 Without soft ramp Acceleration time 28 s2
9 s29 s
Deceler
ation time 22 s23 s23 s
With soft ramp Acceleration time 36 s36 s36 s
Deceleration time 35 s35 s35 s
F-45-48-PCR Without soft ramp Acceleration time 11 s12 s12 s
Deceleration time 12 s13 s13 s
With soft ramp Acceleration time 22 s22 s22 s
Deceleration time 22 s22 s22 s
S-24-11-AT Without soft ramp Acceleration time 18 s18 s18 s
Deceleration time 17 s17 s17 s
With soft ramp Acceleration time 29 s29 s29 s
Deceleration time 30 s30 s30 s