2.4.6 Rotor F-45-48-5-PCR
2.4.7 Rotor S-24-11-AT
This rotor is only intended for use with 1.5 mL/2.0 mL tubes.
The following tubes must not be used in this rotor:
• Adapters for 0.2 mL, 0.4 mL, 0.5 mL and 0.6 mL tubes and the corresponding tubes.
• Spin columns
Rotor F-45-48-5-PCR
Fixed-angle rotor for 48 PCR
tubes
Max. g-force: 11 710 × g
Max. speed: 10 500 rpm
Max. load (tube and
contents):
48 × 0.43 g
Tubes Tube Adapter Adapter base Max. g-force:
Capacity Tube diameter Max. speed
Tubes per adapter/
rotor
Centrifugation radius
– conical 11 710 × g
0.2 mL Ø 6 mm 10 500 rpm
–/48 9.5 cm
PCR strips – conical 11 710 × g
0.2 mL Ø 6 mm 10 500 rpm
–/6 × 8 9.5 cm
Rotor S-24-11-AT Bucket for tubes:
4 × 1.5 mL/2.0 mL
Max. g-force: 16 049 × g
Max. speed: 12 700 rpm
Max. load per bucket (tube
and contents):
4 × 3.75 g
Vessels Vessel Adapters Adapter base Max. g-force:
Capacity Order no.
(international)
Tube diameter Max. speed
Vessels per
adapter/rotor
Centrifugation radius
Vessel – – 16 049 × g
1.5 mL/2.0 mL Ø 11 mm 12 700 rpm
-/24 8.9 cm