New Brunswick Operating manual
3. The SET and MAINT INDICATORS will flash and the accumulated running time will be
displayed in hundreds of hours (i.e., “02” equals 200 hours; “102” equals 10,200
hours). This display will continue for 10 seconds and then default to the previous
mode readout.
6.9 Maint Indicator
After 10,000 hours of operation, the MAINT INDICATOR will illuminate. Preventive
maintenance is recommended at this point.
To deactivate the MAINT INDICATOR, follow these steps:
1. Press SELECT until the HRS INDICATOR is illuminated.
2. Simultaneously press the σ and τ KEYS.
3. Press the τ KEY.
6.10 Automatic Restart
In the event of a power failure, the C-76 is equipped with an automatic restart function.
If the shaker was in operation at the time of the power interruption, it will begin to
operate at its last entered setpoint when the power is restored. The
LED DISPLAY will
flash, indicating that a power failure has occurred. Press any key to stop the flashing.