Manual
25
New Brunswick Operating manual
6.4 Timed Functions
The shaker can be programmed to stop automatically after a preset time period of 0.1
hour to 99.9 hours. There must be power to the shaker in order to set the timer; however,
a timed run can be initiated while the unit is either shaking or stopped.
6.4.1 Setting the Timer
To set the time:
1. Press the SELECT KEY until the HRS INDICATOR is illuminated.
2. Press either σ or
τ
KEY to enter the SET MODE and set between 0.1 - 99.9
hours.
3. While the SET INDICATOR is illuminated, press the START/STOP KEY to
program the time (and start the run). The TIME INDICATOR will light and
remain on for the duration of the run. At the end of the timed run, the display
will read
OFF, and the TIME INDICATOR will flash.
1. To disable the alarm (flashing
TIME INDICATOR), press the SELECT KEY and
change to any other function.
6.4.2 Cancelling the Timer
To cancel the timer without stopping the shaker:
1. Press the SELECT KEY until the HRS INDICATOR is illuminated.
2. Press either σ or
τ
KEY to enter the SET MODE and set between 0.1 - 99.9
hours, then immediately press the START/STOP KEY. The TIME INDICATOR
will cease to flash and the display will read OFF.
CAUTION!
Holding the
σ
or
τ
key for more than 0.5 seconds will cause
the speed setpoint to change. Should this occur, resetting
will be necessary.