8.3.2 Checking the thermal module
• Carry out an inspection at approx. 20 °C.
• Make sure that the cuvette shaft is empty for the check.
1. Select the Temperature unit function in the
Device calibration group and confirm with enter.
2. Make sure the cuvette holder is empty, and start
the test procedure with [Next >].
The following temperature control test at
5 temperatures takes approx. 30 minutes. The
estimated remaining time is displayed during the
test.
3. The results are displayed after the test has been
completed.
A positive test result is indicated by "OK" behind
the temperature value. If the result "failed" is
displayed for at least one temperature please
contact Eppendorf Service.