8.3.1.1 Checking photometric accuracy
1. Select the Spectrometer unit function in the
Device calibration group and confirm with enter.
2. Select whether you want to check the wavelength
systematic error or photometric accuracy, and
confirm with enter. Press [Next >] to switch to
the measurement.
3. Follow the instructions on the device display and
start by measuring the A0 blank filter, and then
the first test filter A1.
The device measures the test filter 15 times at
9 wavelengths, and then shows the average value
and the VC values for the series of measurements
for all 9 wavelengths.