Examples:
λ grid: 100 nm, min. Δ Abs: 0.050:
The peak is not detected because the λ grid is too
large: The absorbance values on the left edge of the
grid are higher than the absorbance of the peak.
λ grid: 20 nm, min. Δ Abs: 0.200:
The peak is not detected because the predetermined
value for min. Δ abs is too high. The difference of the
absorbance of the peak and the lowest absorbance in
the grid is less than 0.2 A.
λ grid: 20 nm, min. Δ Abs: 0.050:
The peak is detected.