8.3.2 Checking the fluorescence unit
8.3.3 Device self-test
The frequency of the automatic self-test (approx. 1 minute) can be set using the Device settings function
(see Device settings on p. 64). The factory setting for Self-test interval is "Weekly".
The self-test checks the following:
Verification of the detector
Determination of the random error across the entire available spectrum
Verification of the light source
Verification of the maximum available energy of the light source and quality of the light transmission
through the device
Determination of the signal random error at the reference sensor
Determination of the signal level at the reference sensor
Separate determination of light intensity in UV range
Determination of the systematic and random error of the wavelength
Position of an intensity peak in the UV range of the spectrum
Precision of the position of an intensity peak in the UV range of the spectrum
1. Select the Fluorescence unit function in the
Device calibration group. Confirm with enter.
2. Place the F1 filter in the cuvette shaft. Press the
Measure softkey.
The device measures the test filter 15 times at 2
emission wavelengths. After the measurement
has been completed, the display shows 2
characteristic values: "Ratio" as the dimension for
the correct adjustment and "CV" as the dimension
for the noise.
3. Compare the characteristic values to the values in
the provided table. If the measured values do not
agree with the permitted range of values, contact
Eppendorf Service.