Owner's manual

Troubleshooting
Eppendorf BioSpectrometer
®
basic
English (EN)
78
Calculation not
possible because of
division by zero.
Absorbance result
or Formula "b"
parameter is zero.
An absorbance result was divided by
a "zero" value during the evaluation
of a Division type method (Dual
wavelength method group). This is
not mathematically permissible.
Check the reagents and samples used
and repeat the measurement.
Do not enter "zero" as a value for the
Formula b parameter.
There is only one
measurement left to
be performed in
this series of
measurement.
The maximum
number of
measurements
within one series of
measurements has
been reached.
The number of measurements in one
measuring series is limited to 99.
Start a new series of measurement
after maximally 99 measurements.
Invalid zoom
interval!
Error in the process results method step
in the Zoom mode.
Permissible zoom range for the
wavelength scale:
Wavelength interval at least 10 nm
Entries for wavelengths only within
the range programmed in the
parameters for the method.
Permissible zoom range for the
absorbance scale:
Absorbance interval at least 0.02 A
Upper and lower limit for absorbance
interval +3 A or –3 A
Please observe the stated limits in the
zoom procedure.
Problem Cause Solution