Manual

82
10 Short instructions
Programming
The factory-set method programs may be changed as required.
Example:
Programming of the unit "
µ
g/mL" and of calculation via standard (500
µ
g/mL) for the Protein method.
PROTEIN
ABSORBANCE
Sample
PROTEIN PAGE 1–3
CALCULATION STD –
FACTOR –
ABSORBANCE
*
WARBURG –
Blank
PROTEIN PAGE 4–4
CUVETTE 10 mm
*
5 mm –
2 mm –
1 mm –
PARAMETER END
Select Protein method
Select calibration
Enter standard concen-
End programming
PROTEIN PAGE 3–4
UNIT mg/mL –
µg/mL
*
STD - - - - - µg/mL
PROTEIN PAGE 1–3
CALCULATION STD –
FACTOR –
ABSORBANCE
*
WARBURG –
PROTEIN PAGE 3–4
UNIT mg/mL
*
µg/mL –
STD - - - - mg/mL
PROTEIN PAGE 2–4
CORR.WITH A
320 OFF *
ON –
STD MEASUREMENT 1 x
*
2 x –
3 x –
Select the unit µµ
µµ
g/mL
Confirm
Enter
4
Protein
Begin programming
Parameter
with standard (STD)
Conversion
Dilution
or
Confirm
Enter
(After confirmation, the display
moves to the next selection block)
tration and confirm
(After confirmation, the display
moves to the next selection block)
0
Sample No.
0
Sample No.
5
OD 600
Enter
Parameter
▲▲
▲▲
Conversion
Dilution
or
or
10_Kurz_e.fm Seite 82 Montag, 20. Februar 2006 11:33 Uhr