8 Error messages, result flagging and help texts
Result flagging
Error texts in the results display
Flagging Explanation
1.586 A
260
Flagging of A
260
in the display or on the printout
(for method "Protein direct" only):
The method was calculated using the Warburg formula.
0.015 A
320
Flagging of A
320
in the display or on the printout
(for method "Protein direct" and for nucleic acid methods only):
The absorbances at 260, 280 and 230 nm are corrected with the absorbance
at 320 nm (see Section 6, "Programming").
Error text Explanation / Cause Solution
+++++ The absorbance measured is greater
than 3.0 A.
Dilute the sample.
Check the cuvette
(height of light path must
be 8.5 mm).
Clean the cuvette shaft
(see Section 9).
Insert the cuvette correctly (the
measuring window must be facing
the light path).
Use a cuvette made of material
that transmits light at the
measuring wavelengths used
(e.g. quartz glass or UVette
®
from
Eppendorf for nucleic acid
measurement).
!
!
!
!
! The calculated result cannot be displayed
(value too high).
Check the parameter
(Is the factor too high?).
– – – – – (Instead of a value for the ratio:)
Ratio cannot be calculated because one of
the absorbance values used for calculating
the ratio is 0 A or > 3.0 A.
Repeat the measurement
(dilute sample if necessary).