5 Operation
5.5 Measuring OD 600
Select method
Measuring procedure
Blank measurements remain stored until the date changes
(for further details, see "Measuring nucleic acids").
Measure blank
Measure sample
Measure
next sample
To measure the next sample, press the key again.
Sample dilution
Sample dilution in the measuring cuvette can be entered
using the key before the measurement begins and
is then included automatically in the result calculation
(see "Measuring diluted samples").
The OD 600 measurement is a stray-light measurement;
the result is therefore heavily dependent on the geometry
of the light path, which may vary between photometers
from different manufacturers.
5
OD 600
OD600
Blank
Sample
or
Blank
OD600 BLANK
0 . 0 0 0 A
Blank
Sample
or
Sample
OD600 SAMPLE 001
0 . 4 3 0 A
Sample
Dilution