If the shaker is stopped:
5. Press the START/STOP KEY. The shaker will start in untimed mode.
6. Press the START/STOP KEY again. The shaker will stop and the display will read
OFF.
7. Press the START/STOP KEY a third time; the TIME INDICATOR will light and the
shaker will start the timed run.
To disable the visual alarm (flashing TIME INDICATOR), press the SELECT KEY and
change to any other function.
To cancel the timer while the shaker is running:
1. Press the SELECT KEY until the HRS indicator lights.
2. Press the KEY until 0.0 is displayed, then press the START/STOP KEY. The display
will read OFF, the shaker will stop, and the TIME INDICATOR light will turn off.
3. Press the START/STOP KEY to continue in untimed mode.
To cancel the timer while the shaker is stopped:
1. Press the KEY until 0.0 is displayed, then press the START/STOP KEY. The TIME
INDICATOR
will light and the shaker will run.
2. Press the START/STOP KEY. The shaker will stop and the TIME INDICATOR will turn
off.
3. Press the START/STOP KEY a third time, and the shaker will run in untimed mode.
6.5 Alarm Functions
In addition to the visual alarm mentioned above, the Excella E-25 and E-25R shakers
have an audible alarm that is activated at predetermined times. It can be deactivated by
using the
MUTE function:
1. Press the
SELECT key until the MUTE indicator illuminates.
2. Press the
(UP) or (DOWN) KEY to display ON, then press the SELECT KEY.
To reactivate the audible alarm:
1. Press the SELECT key until the MUTE indicator illuminates.
2. Press the (UP) or (DOWN) KEY to display OFF, then press the SELECT KEY.