User Manual

Technical data
Eppendorf
®
PlateReader AF2200
English (EN)
62
9.5 Specifications
9.6 Fluorescence intensity
Use Universal lab device
Interface USB
Measurement Software-controlled
Plate definitions Via software
Temperature control 5 °C above ambient temperature up to 42 °C
Microplates 6-well to 384-well plates (ANSI/SLAS standard
formats)
Shaking the plates Linear and orbital shaking, amplitude adjustable
from 1 to 6 mm in 0.5 mm increments
Filter handling Manually replace the filter outside of the device.
Optics Quartz glass lens
Light source High-performance Xenon flash lamp,
service life: 10
8
flashes
Fluorescence detector Photomultiplier (PMT)
Absorbance detector Silicon photodiode
Noise level < 60 dBA
Parameter Characteristics
Filter Predefined Eppendorf filter slide or filter slides with
user-specific filters.
Wavelength range Excitation: 230 – 600 nm
Emission: 330 – 600 nm
Setting the signal
amplification (= gain)
Values Measuring range
Manual 1 - 255 0 - 60,000 RFU
Optimal Automatic 0 - 60,000 RFU
Calculated from the well Automatic 0 - 60,000 RFU
Zeit einstellen Characteristics
Integration time 20 - 2000 µs
Delay time 0 - 2000 µs