8.6.1 Test criteria
All specifications can be changed without prior notification.
Microplate specification Reliable range of values
Upper fluorescence sensitivity < 20 pM fluorescein
Upper fluorescence uniformity < 3% CV
Upper fluorescence precision < 2% CV
Lower fluorescence sensitivity < 100 pM fluorescein
Lower fluorescence uniformity < 3% CV
Lower fluorescence precision < 2% CV
Absorbance accuracy 0 – 2 OD: ≤ ±1% + 10 mOD
2 – 3 OD: ≤ ±2.5%
Absorbance baseline stability (1 sigma) < ±10 mOD