User Manual

Inserting the filter slide
Eppendorf
®
PlateReader AF2200
English (EN)
33
2. Adjust the filter definitions as follows:
The Overview tab and four tabs for the filter definition are shown in the Filter definitions dialog window.
Overview: The definition of the current filter slide is displayed for the user in the overview.
Description of the filter slide: Enter the description of the filter slide or the description will be
automatically generated.
Position 1 - 4: Editor for the filter definition of filter pairs on positions 1, 2, 3, 4.
Select the corresponding filter position and enter the new wavelength, bandwidth and the new
measuring mode for every new filter.
Measuring mode: In the drop-down list, select "FI" for fluorescence intensity, "ABS" for absorbance
and "Empty" for positions without filters.
Special characters (spaces, ?, $, %, ., / etc.) are not allowed in the filter slide description
(exception: '_').
The filter slide description is part of the filter slide code. Each filter slide must, therefore,
receive a unique name.