6.2.7 Cell viability (CellTiter-Blue Assay)
Introduction
The CellTiter-Blue Cell Viability Assay provides a homogeneous, fluorometric method for monitoring cell
viability. The assay is based on the ability of living cells to convert a redox dye (resazurin) into a fluorescent
end product (resorufin). Nonviable cells rapidly lose metabolic capacity and thus do not generate a
fluorescent signal.
Default measurement parameters
Excitation wavelength: 535 (25) nm
Emission wavelength: 595 (35) nm
•Number of flashes: 25
Settle time: 0 ms
•Mode: Top
Lag time: 0 μs
Integration: 20 μs
Gain: "optimal"
Plate definition: Eppendorf Cell Culture Plate 96/F, black/clear" or comparable plates
User defined method parameters
The measurement is performed at 535 / 595 nm with default measurement parameters.
Define the number of blank replicates (possible number: 1 – 6).
Define the number of control replicates (possible number: 1 – 6).
Define number of samples and sample replicates (max. 58, depends on number of blanks and controls).
Define appropriate sample identifier (default settings are ID_1, ID_2, ..).
Check if entered parameters are correct.
Required plate layout: Outer wells are left empty by default. Pipette in a vertical order, starting with
blanks at position B2, controls are pipetted in subsequent wells, samples must be placed in the
following wells.
Press Start.
For method results and data evaluation principle see chapter method results and evaluation procedure.
The method is restricted to 96-well microplates.
Because of faster evaporation the wells lying on the outer sides of the plate are not included in
the evaluation.
Please fill them with buffer or distilled water.