User Manual

Predefined methods
Eppendorf
®
PlateReader AF2200
English (EN)
62
User defined method parameters
Using the predefined UV/Vis filter slide the measurement is performed at 600 nm with default
measurement parameters.
Define the number of blank replicates (possible number: 1 – 8).
Define the number of standards (possible number: 2 – 12).
Define number of standard replicates (possible number: 1 – 8).
Define the concentrations of each standard (Press triangle button to open the menu).
Define the number of samples (max. 93, depends on number of blanks and standards).
Define the unit of your concentration.
Check if entered parameters are correct.
Required plate layout: Pipette in a vertical order, starting with blanks at position A1. Standards are
pipetted in subsequent wells, samples must be placed in the following wells.
Press Start.
For method results and data evaluation principle see chapter method results and evaluation procedure.
Plate layout
Abb. 6-5: Metho d strip: Protein quantification (BCA 562 (600) nm with standards)
Fig. 6-5: Method strip: Protein quantification (BCA 562 (600) nm with standards)
The selected number of blanks, standards and samples are shown in the plate view.