5.4 Defining Kinetic Measurements
The following example describes a kinetic measurement of a 96 well plate.
Select the 96 well plate (e.g. Eppendorf 96 Flat Clear UV-VIS) from the Plate definition drop-down list,
and select all wells in the Pattern program element.
Double-click the Kinetic Cycle program element and define as follows:
Cycles/Number of cycles: 50
Kinetic Interval (intervals between measurements): Select Use kinetic interval and enter: 2 minutes
30 seconds.
Double-click the Absorbance program element and define as follows:
Wavelength/Measurement: 462 nm
Read/Number of reads: 25