5.2 Defining Endpoint Measurements
The following example describes an Absorbance Endpoint Measurement in all wells of a 96 well plate.
Select a 96 well plate (e.g. Eppendorf 96 Flat Clear UV-VIS) from the Plate definition drop-down list. If
the Pattern program element is not visible, click the link Pattern. It is recommended to use the Pattern
program element in every workflow, even if all wells are measured.
Double-click the Absorbance program element from the Explorer bar, and define the Workflow as
follows:
Wavelength/Measurement: 462 nm
•Read/Number of reads/flashes: 25 (per well)
Settle time (time between moving the plate and starting the measurement): 0 ms