PlateReader AF2200
English (EN)
29
4.5.5 Settings Menu
Heating This command is used to set the target temperature of the
instrument manually.
Select or enter the Target temperature and click Set and On to
start instrument heating.
•Click the Read button to display the current temperature inside
the instrument or click the Auto check box to have it read
automatically.
•Click Off to stop heating.
Click the down button to display the heating graph and
click the up button to hide it.
Click the close button to exit the Heating dialog box.
Properties Select Properties to set a new alias name for the instrument.
Enter a new name in the New Alias field and click Set Alias to
confirm.
These settings take effect after restarting the software.
Filter Definitions
Measurement Mode Select the appropriate filter position and enter the new wavelength,
bandwidth, and measurement mode for each new filter.
Choose from the dropdown list:
FI for fluorescence intensity,
ABS for absorbance measurements, and
Empty for filter-free positions.
Wavelength Enter the filter wavelength.
For fluorescence intensity measurements, set the filter wavelength
within the allowed range of the instrument.
Absorbance filters are definable between 230 and 1000 nm
(Excitation only).
Bandwidth Enter the bandwidth (nm) of the filter.
Description This field can be used for individual user remarks about the filter,
e.g. filter name, application, etc.