Owner's manual
9 Technical data
Centrifuge 5424 / 5424 R — Operating manual
34
9 Technical data
Operating manual — Centrifuge 5424 / 5424 Rseeon p.Fig.Tab.p.
9.1 Power supply
Centrifuge 5424
Centrifuge 5424 R
9.2 Ambient conditions
Mains power connection 230 V, 50 to 60 Hz
120 V, 50 to 60 Hz
100 V, 50 to 60 Hz
Current consumption: 1.9 A (230 V)
3.8 A (120 V)
4.0 A (100 V)
Power consumption: max. 250 W
EMC: Interference emission (radio
interference)
EN 61326 - category B
EMC: Noise immunity EN 61326
Overvoltage category: II
Fuses: 3.15 AT (230 V)
6.3 AT (120 V / 100 V)
Mains power connection 230 V, 50 to 60 Hz
120 V, 50 to 60 Hz
100 V, 50 to 60 Hz
Current consumption: 1.6 A (230 V)
3.2 A (120 V)
3.4 A (100 V)
Power consumption: max. 350 W
EMC: Interference emission (radio
interference)
EN 61326 - category B
EMC: Noise immunity EN 61326
Overvoltage category: II
Fuses: 3.15 AT (230 V)
6.3 AT (120 V / 100 V)
Environment: For indoor use only.
Ambient temperature: Centrifuge 5424: 2 to 40°C
Centrifuge 5424 R: 10 to 40°C
Max. relative humidity: 10 to 75 %, non-condensing humidity
Atmospheric pressure: Use up to an altitude of 2000 m above MSL.
Degree of contamination: 2