User Manual
35
Innova 3100
M1231-0050 Operating manual
6
6
S
S
P
P
E
E
C
C
I
I
F
F
I
I
C
C
A
A
T
T
I
I
O
O
N
N
S
S
This chapter provides technical details of interest, but not necessarily
essential for operation of the instrument.
SHAKING
Speed
25-400 RPM
Motion
½ inch (12.7 mm) diameter circular orbit
Indication
LED digital electric display, 1 RPM increments
Setpoint & Control
Digital adjustment with PI microprocessor control and
instantaneous visual feedback
Accuracy
±
1 RPM
DRIVE
Triple eccentric counterbalanced ball bearing drive.
TEMPERATURE
Range
5.0
°
C above ambient to 80
°
C
Range with Cooling
Option
5.0°C above coolant temperature to 80°C
Control Stability
± 0.1°C by use of PI microprocessor controller and pulse width
modulation of heater
Setpoint
Digital adjustment increments (0.1°C increments) with
instantaneous visual feedback
Accuracy
± 0.1° @ 30-40°C range
±
0.50
°
C for remaining range
Uniformity
Better than 0.5
°
C
Safety
Heater shuts off if temperature exceeds operating range. Two
safeties provided.
Heater
Low watt density resistance mat heater
TIMER (Shaker)
• Programmable shaking periods from 0.1 hour to 99.9 hours by a digital timer that shuts
off at the end of period and energizes status light.
• Timer counts down and digital display indicates remaining time. Can be deactivated for
continuous operation.
•
Additionally, unit will display total accumulated running time for service information.
AMBIENT OPERATING ENVIRONMENT
0
°
- 40
°
C, 90% humidity, non-condensing
SELF-DIAGNOSTIC STATUS
Warning signal (audible and visible) indicates when shaking speed deviates more than 5
RPM or the temperature deviates more than 1.0°C from setpoint and when timer operation
has expired. The audible alarm can be deactivated/activated by the operator.
REMOTE MONITORING
• Chart recorder outputs for speed and temperature (0-5V).
•
1V per 100 RPM; 1V per 20
°
C, accuracy
±
25mV.