User Manual

35
Innova 3100
M1231-0050 Operating manual
6
6
S
S
P
P
E
E
C
C
I
I
F
F
I
I
C
C
A
A
T
T
I
I
O
O
N
N
S
S
This chapter provides technical details of interest, but not necessarily
essential for operation of the instrument.
SHAKING
Speed
25-400 RPM
Motion
½ inch (12.7 mm) diameter circular orbit
Indication
LED digital electric display, 1 RPM increments
Setpoint & Control
Digital adjustment with PI microprocessor control and
instantaneous visual feedback
Accuracy
±
1 RPM
DRIVE
Triple eccentric counterbalanced ball bearing drive.
TEMPERATURE
Range
5.0
°
C above ambient to 80
°
C
Range with Cooling
Option
5.0°C above coolant temperature to 80°C
Control Stability
± 0.1°C by use of PI microprocessor controller and pulse width
modulation of heater
Setpoint
Digital adjustment increments (0.1°C increments) with
instantaneous visual feedback
Accuracy
± 0.1° @ 30-40°C range
±
0.50
°
C for remaining range
Uniformity
Better than 0.5
°
C
Safety
Heater shuts off if temperature exceeds operating range. Two
safeties provided.
Heater
Low watt density resistance mat heater
TIMER (Shaker)
Programmable shaking periods from 0.1 hour to 99.9 hours by a digital timer that shuts
off at the end of period and energizes status light.
Timer counts down and digital display indicates remaining time. Can be deactivated for
continuous operation.
Additionally, unit will display total accumulated running time for service information.
AMBIENT OPERATING ENVIRONMENT
0
°
- 40
°
C, 90% humidity, non-condensing
SELF-DIAGNOSTIC STATUS
Warning signal (audible and visible) indicates when shaking speed deviates more than 5
RPM or the temperature deviates more than 1.0°C from setpoint and when timer operation
has expired. The audible alarm can be deactivated/activated by the operator.
REMOTE MONITORING
Chart recorder outputs for speed and temperature (0-5V).
1V per 100 RPM; 1V per 20
°
C, accuracy
±
25mV.