8 Technical Specification
Measuring modes : Active and passive
Metering range : 0 to 5.5 ‰ BAC
Accuracy
Range 0 to 1‰
: ± 0.05 ‰ absolute
Range 1 to 2‰
: ± 5 % of measured value
Range > 2‰
: ± 10 % of measured value
Ambient conditions
Temperature range – operation:
: -5 °C to +50 °C
Temperature range – storage:
: -20 °C to +60 °C
Optimum storage temperature
(sensor):
: 0 °C to +25 °C
Ambient pressure:
: 600 to 1400 hPa
Ambient humidity
: 20 to 98 % r.h.
Minimum expiration volume:
:
1,2 liters
Batteries : 4 x LR6 (AA) Alkaline Mignon cells
(Recommended: Duracell
®
Procell
®
)
Number of measurements per battery
pack
: more than 1600 tests, depending on the
batteries and the ambient conditions
Device interface : RS 232 (serial interface). Data can be
transferred to a PC
Dimensions ( L x W x H) : 190 mm x 61 mm x 38 mm
Weight : 275 g incl. batteries
Miscellaneous
Ready after switching on : approx. 6 s
Ready after 0 ‰ measurement : approx. 2 s
Ready after 1 ‰ measurement : approx. 5 s
Sensor type : E100 electrochemical measuring cell
Cross sensitivities : In the form of other substances accompanying
respiration, such as acetone etc. are negligible
Units : configurable: ‰, mg/l
Input : Three-key operation
Power-saving mode : Device switches off automatically after 2.5
minutes of non-operation,the background
lighting switches off after 15 seconds
Display : Background illuminated, fully graphic
Data memory : Stores 9999 measured values with date and time
Mouthpieces : Disposable mouthpiece EnviteC
Software : Optional PC software for statistical analysis, data
transfer and logging
Calibration : Every 6 months
Safety-related control : Every 6 months
Subject to technical modifications!