Dok.-Nr.: 059-07-11000300-k.doc / 07.08 33
Error indication Possible cause Remedy
Device errors which endanger the measuring process (red LED)
Error temperature
Temperature
The temperature
established in the device is
beyond the specified range
of between -5 and +50 °C.
Measurement may be faulty.
Beyond the specified temperature range
the measured values established may
be subject to greater error tolerance
than specified. The device must be
switched off and brought to within the
specified temperature range (by cooling
or heating).
Then the device must be switched on
again.
If the error occurs although the device is
evidently within the specified
temperature range, please contact
Service.
Error pump
Pump
An error was detected in
the sampling system.
Switching the device on and off.
If the fault occurs again during the self-
test or when measuring, please inform
Service.
A faulty measurement cannot occur
because the measured value is rejected
in the case of “Faulty sample collection”
Error memory
Memory
An error has occurred
during reading or writing
access to the internal
memory.
Measuring is not possible.
The measured value may not be
displayed correctly.
Data cannot be read correctly off the
memory and the established measured
values cannot be logged. Measure
again.
If the error recurs, please contact
Service.
Error alcohol sensor
Alcohol sensor
An error was detected on
the alcohol sensor.
Measuring is not possible.
Switch the device off and then on again.
If the error recurs, please contact
Service.
Error calibration
Calibration
An incorrect calibration
date was found.
Measuring is not possible.
Device must be calibrated again, please
contact Service.
If the device has already been
calibrated, it is possible that the current
date settings are not correct.
Here, too, please contact Service.