6.5.4 Storing a Calibration File
Once all of the calibration gases have been calibrated the user can select the option to Store Calibration File”. This
feature will generate a .csv file with all of the calibration information generated during the calibration activity.
If the user has failed to calibrate all of the parameters a warning screen will appear and will indicate “Not all calibrated”.
This is a warning box so that the user can go back and calibrate the remaining parameters, if desired. If the user taps the
“OK” box in the warning message it will return the screen back to the “Calibration Menu” so that the remaining
parameters can be calibrated. If the user does not calibrate the remaining parameters and taps the “Store Calibration