P a g e | 47
2) On the “Pressure Calibration” screen allow the pressures in the 5 pressure blocks to stabilize.
3) Ensure that the sample fittings are disconnected from the wellhead and shielded from wind
4) Tap the “Zero Transducers” button on the lower center part of the touch screen.