10) Now that the Trimble handheld computer has returned to the “Calibration Menu” the user will notice that the
parameter that has been calibrated is now highlighted in green.
11) The user may now select the next gas to calibrate
6.5.3 Pressure Sensor Calibration
The Envision® gas analyzer has 5 internal pressure sensors. These sensors should be field calibrated at the beginning and
periodically during the monitoring event. The pressure sensors may be zeroed from two different locations in the Elkins
Earthworks® Analyzer software.
From the “Calibration Menu”
1) Tap the “Pressure Zero” button. The screen will change to the “Pressure Calibration” screen