The “Meter Status” screen gives the user useful information about the status of the Envision® meter.
Battery time remaining Indicates approximately how much time the Envision® has until the battery
needs charged.
Meter Temperature This displays the internal temperature of the Envision® gas analyzer.
O
2
cell replacement The date indicated is the date the O
2
cell was installed.
Factory calibration The date indicated is the last date the Envision® gas analyzer was last calibrated at
the factory.
Firmware version Indicates the firmware version installed in the Envision® gas analyzer.
Heater - Indicates when the heater inside the Envision® would turn on if the external heater battery
pack is connected.
Meter Error Conditions Will display the last 10 meter error conditions. This is useful when talking to
tech support staff at Elkins Earthworks®.