Manual

800.745.8218 / 843.576.1147 / www.environmentalexpress.com Environmental Express • 4
HotBlock™ Pro™: Operation and Instruction Manual Part Numbers / Model Information / Specications
Above, the 35-well HotBlock™ Pro™ for 100mL samples is shown with the external
controller. The separate controller is designed to work with one or two blocks. See
individual block specications. For digestion vessels and accessories, see pages 13-16.
Specications for SC182:
Block Footprint: 21.5”w x 15”d x 6.5”h
Block Weight: 54 lb.
Sample Capacity/Sample Size: SC182 - 35 sample, 50mL
Electrical: 110V*, 9A per block
Block Temperature Range: Ambient to 180°C
Block Temperature Resolution: 0.1°C
Block Temperature Uniformity: ± 0.2°C (measured across block)
Controller Dimensions: 12”w x 10.5”d x 8”h
Controller Weight: 13 lb.
Internal Thermocouple: Type K
External Thermocouple: Records sample temperature
Part Number SC183
HotBlock™ Pro™ SC182
HotBlock™ Pro™ Controller SC180