Specications for SC181:
Block Footprint: 21.5”w x 15”d x 6.5”h
Block Weight: 58 lb.
Sample Capacity/Sample Size: SC181 - 54 sample, 50mL
Electrical: 110V, 9A per block
Block Temperature Range: Ambient to 180°C
Block Temperature Resolution: 0.1°C
Block Temperature Uniformity: ± 0.2°C (measured across block)
Controller Dimensions: 12”w x 10.5”d x 8”h
Controller Weight: 13 lb.
Internal Thermocouple: Type K
External Thermocouple: Records sample temperature
Part Number SC183
Above, the 54-well HotBlock™ Pro™ for 50mL samples is shown with the external controller.
The optional external thermocouple (shown left, in a specially designed holder) can be used
to monitor and record sample temperatures.
The separate controller is designed to work with one or two blocks. See individual block
specications. For digestion vessels and accessories, see pages 13-16.