Manual

SERVICE TAB
Maintenance
This section is designed to remind the operator when it is time for
replacement of parts or calibration of processes.
Failure to adhere to AutoBlock maintenance instructions and/or
requirements can void all warranties and result in damage to the
A
utoBlock, loss of samples and/or losses not covered by the
warranty. Please review the warranty section of this manual.
After selecting the MAINTENENCE button, there are several items to
be addressed on (3) pages. Next to each maintenance item shows:
date last serviced - service indicator - next service date
Maintenance items that need addressing are:
Graphite Calibration Teflon Tubing Pump 1 Teflon Tubing Pump 2
System Clean Pump Tubing 1 Pump Tubing 2
VICI Emitters Pump Calibration 1 Pump Calibration 2
HEPA Filter Check Waste Drum
After service has been completed double click the SERVICE button to update
the screen. This will change the button from red to green and change the service dates.
Select MAIN MENU button to continue back to the SERVICE TAB screen.
By selecting the Main Menu button, the new information has been saved
before continuing onto the next/previous page.
manual screens
this third screen is only shown in a
dual system
Options
AutoBlock comes with 2-pages of recommended
preset features. Recommended service days may be
changed by the operator to increase or decrease based
on usage and standard laboratory protocol.
Time entered is in days.
Options included:
By selecting the UP or DOWN button
each item will change its current value
Recommended service days have already been provided.
Automatic Reminders service (days)
System Clean 30
HEPA Filter 180
VICI Emitters 365
Graphite Calibration 365
Fan Speed 100%
Pump Tubing 365
Teflon Tubing 365
Pump Calibration 30
Waste Drum 30
Select MAIN MENU button to save data & continue back to the SERVICE TAB screen.
365
100
options screen
365
ENVIRONMENTAL EXPRESS
For technical support call 1-800-745-8218 or send email to INFO@ENVEXP.COM
5588SERVICE rev4 30 9/15/09