Unit installation

MAC 10
®
LEDC
4x4 Fan Filter Unit
12 Innovators in Clean Air Technology | www.envirco.com | Operation & Maintenance
Installation & Service Manual
Troubleshooting
Mode Choice:
Verify mode setting choice to DIP switch S1 (Control Mode), which is manual mode and then retry.
Motor Issues in Manual Mode:
(If you are in a network or analog mode, contact your controls contractor for troubleshooting assistance; if you continue to
need assistance, contact the factory.)
Unit is not adjustable:
Step 1. Verify that rotation of the manual speed potentiometer does not change the RPM. If rotating does nothing, remove
the electrical box cover, then remove the 4-pin motor connector from the control board and reinstall 180 degrees rotated.
Also insure that the 4-pin connection wires are pushed down securely onto connector pins. Again adjust the knob to verify
proper fan operation. Note: The 4 pin connector is on the 1/8” white cable from motor.
Low Air Velocity:
Step 1. Check to be sure that the manual speed potentiometer is set correctly.
Step 2. Check incoming power supply for proper voltage (120,208-240,277/24).
Step 3. Examine the HEPA lter.
High Air Velocity:
Step 1. Check to be sure that the air ow settings are correct.
Filter Issues:
Non-Laminar Flow and/or Excessive Contamination:
Step 1. Insure that no large obstructions are upstream of airow pattern.
Step 2. Determine that no other air-moving devices are operating in or around clean room which disrupt room’s
airow pattern.
Step 3. Check air velocity and if low, conduct the “Low Air Velocity” procedure outlined above.
If failure continues, contact Envirco Technical Support at 800-884-0002.
Net LED Status Denition
Green LED OFF Power Lost or No Communications
Green LED Flickering Network Data Trafc In Progress
Green LED ON A/B network wires are swapped
Red Network LED Green Status LED