Specifications
20
Troubleshooting
2. Check that the FREEZE button can clearly be depressed.
A distinct click feeling should be felt.
Video quality is different on the TV than on the unit:
Amigo is optimized for the built in LCD viewing display. TV
display may differ slightly.
The Battery does not charge:
Make sure the Amigo Power Supply is used with the unit.
Check and see if the battery is properly inserted with the
contacts out of the housing. The LED should be blinking if
the battery is not full. AMIGO will recharge depleted
batteries or partially depleted. Fully charged batteries are
not recharged until their level is low enough.
Amigo does not remember the last settings used:
AMIGO will remember the last settings used when the
power is shut OFF with the POWER button. If the battery
runs too low and the Amigo shuts off then the unit will not
save the last settings into memory.
Picture is not moving:
Check to make sure the FREEZE button is not stuck. A click
feeling should be clearly felt when depressed. The FREEZE
button must be pressed for at least 1 second.
LCD Swivel is loose and closes on unit:
If this ever happens, use a small Philips screwdriver to
adjust the swivel tension. (see page 13)