11 Tap the side of the
reservoir to remove any air
bubbles.
12 To purge air bubbles that
have risen to the top of the
reservoir, push up on the plunger
until you see insulin in the
tubing.
13 Without pulling, turn the
plunger counter-clockwise to
remove it from the reservoir.
Changing your infusion set
Removing the reservoir
Each time you remove and replace a reservoir in your pump, you have to rewind your pump, and fill
the infusion set with insulin.
1 Remove the entire infusion set from your body.
2 If attached, remove the activity guard.
3 Turn the tubing connector half-turn counter
clockwise, then pull the reservoir and connector
out from the pump.
tubing connector
half-turn counter-clockwise
4
Safely dispose of the used reservoir and infusion set item in a sharp's container.
5
You must now rewind your pump as described in the next section.